Cold chain: difference between revisions

Diff·revision 9 → 10·16:04, 10 Sep 2024

Difference between revision 9 and revision 10 of Cold chain. 16 lines changed; the page grew by 1,701 bytes.

Revision 9 — 08:45, 26 Aug 2024
CDMO_Caradoc (talk)
add the data-logger point: an unlogged shipment has no excursion record
8,537 bytes +927
Revision 10 — 16:04, 10 Sep 2024
CounterfeitCato (talk)
give the vial sizes in millilitres of nominal capacity
10,238 bytes +1,701
1{{Infobox concept1{{Infobox concept
2| name = Cold chain2| name = Cold chain
+3| image = cold-chain.svg
+4| caption = A cold chain is a series of custody transfers, each of which is a candidate point of failure and each of which is expected to leave a temperature record.
3| Scope = Manufacture, warehousing, freight, last mile, end use5| Scope = Manufacture, warehousing, freight, last mile, end use
4| Principal controlled variable = Temperature6| Principal controlled variable = Temperature
51; Chemical threshold indicators : Change colour irreversibly when a threshold is crossed. Freeze indicators of this type are the standard means of detecting a freezing excursion in vaccine distribution.53; Chemical threshold indicators : Change colour irreversibly when a threshold is crossed. Freeze indicators of this type are the standard means of detecting a freezing excursion in vaccine distribution.
52; Time–temperature integrating indicators : Accumulate a colour change as a function of both temperature and time, approximating the cumulative thermal exposure that matters for degradation.54; Time–temperature integrating indicators : Accumulate a colour change as a function of both temperature and time, approximating the cumulative thermal exposure that matters for degradation.
+55
+56|+ What each monitoring device can and cannot establish
+57| !Device | Establishes | Cannot establish |
+58|---|---|---|
+59| Electronic data logger | Full time–temperature profile; permits mean kinetic temperature | Product temperature, unless placed in the payload |
+60| Minimum–maximum thermometer | Extremes reached | Duration at any temperature |
+61| Freeze indicator | That a freezing threshold was crossed | By how much, or for how long |
+62| Time–temperature integrator | Cumulative exposure above a reference | The shape of the profile |
+63| Vaccine vial monitor | Cumulative heat exposure at the individual vial | Freezing exposure |
+64
+65=== Calibration and traceability ===
+66A temperature record is only as good as the calibration of the instrument that produced it. Compendial guidance on monitoring devices expects calibration traceable to a national metrology institute, at intervals appropriate to the device, and at points within the range of interest rather than only at ambient temperature. A logger calibrated at 25 °C and used at 5 °C may carry an error larger than its stated accuracy.{{r|usp1118}}
+67
+68The conventional expectation for a logger used in 2–8 °C distribution is an accuracy of ±0.5 °C across that range. This is an expectation drawn from qualification practice and from performance specifications for prequalified devices; it is not a requirement imposed by a general chapter, and articles asserting it as a standard have been corrected on this point.{{r|whotrs961,usp1118}}
5369
54== References ==70== References ==